yDO!BOOKEƒy[ƒWƒŠƒ“ƒNz fi-04@@@110 / 228 10•bŒã‚ÉBOOK‚̃y[ƒW‚Ɉړ®‚µ‚Ü‚· |
|
Azbil Corporation No. SS2-MGS200-0100 - 3 - PERFORMACE SPECIFICATION Accuracy (in combination with the model MGG14C converter) Vs = Velocity of setting range *)High accuracy calibration is available for sizes of 40 to 125mm (1. to 5 inches). Reference Conditions for high accuracy calibration: . Fluid: Tap water . Fluid temperature:15 to 30‹C . Ambient temperature:15 to 30‹C . Ambient humidity:30 to 70% . Length of straight pipe: Upstream>20~DN Downstream>5~DN (DN: Pipe diameter nominal) . Properly grounded . Properly centered Accuracy is guaranteed by the totalized flow volume under the condition of continuous flow measurement for 30 seconds or longer. PHYSICAL SPECIFICATIONS Main body material Measuring pipe materials SUS304 stainless steel Case SUS304 stainless steel Terminal box Aluminum alloy (remote model) Corrosion-proof baked epoxy paint Color Light beige (Munsell 4Y7.2/1.3) Process wetted materials Lining PFA (diameter 15 to 125 mm) Electrode SUS316L, ASTM B574 (Hastelloy C-276 equivalent), Titanium Ferrule for welding SUS304, SUS316L Gasket EPDM (Ethylene propylene diene monomer) Silicone rubber Structure of electrode Internal insertion (electrode can not be removed) Vs (m/s) Velocity during measurement >Vs ~ 20% Standard Accuracy (Calibration Code A) *)High Accuracy (Calibration Code U) 1.0 < Vs < 10 }0.5% of rate }0.35% of rate 0.1 < Vs < 1.0 }(0.1/Vs+0.4)% of rate }(0.1/Vs+0.25)% of rate Vs (m/s) Velocity during measurement < Vs ~ 20% Standard Accuracy (Calibration Code A) *)High Accuracy (Calibration Code U) 1.0 < Vs < 10 }0.1% of Vs }0.07% of FS 0.1 < Vs < 1.0 }0.2(0.1/ Vs+0.4)% of Vs }0.2(0.1/Vs+0.25)% of Vs 4 9 |