【DO!BOOK・ページリンク】 fi-04 141 / 228 10秒後にBOOKのページに移動します |
|
No. SS2-MTG300-0300 Azbil Corporation - 4 - Velocity V (m/s) = K × Q K = Conversion factor = 1/3600 × 4/(..D2) × 10002 D = Size (mm) Q = Flow rate (m3/h) Temperature range and pressure range of process fluid Refer to the following. Size: 2.5 to 10 mm (0.1 to 3/8 inch) Size: 15 to 200 mm (1/2 to 8 inches) Measurable electrical conductivity 10 ..S/cm or greater Measurement flow range Size Maximum flow velocity range is 0 to 0.3 m/s (0 to 0.98 ft/s) Maximum flow velocity range is 0 to 10 m/s (0 to 32.8 ft/s) Conversion Minimum range Maximum range factor K mm inches m3/h GPM m3/h GPM 2.5 0.1 0 to 0.00531 0 to 0.02335 0 to 0.1767 0 to 0.778 56.59 5 0.2 0 to 0.02121 0 to 0.09337 0 to 0.7068 0 to 3.112 14.15 10 3/8 0 to 0.08483 0 to 0.3735 0 to 2.827 0 to 12.44 3.537 15 1/2 0 to 0.1909 0 to 0.8404 0 to 6.361 0 to 28.00 1.572 25 1 0 to 0.5302 0 to 2.335 0 to 17.67 0 to 77.80 0.5659 40 1. 0 to 1.358 0 to 5.976 0 to 45.23 0 to 199.1 0.2210 50 2 0 to 2.121 0 to 9.337 0 to 70.68 0 to 311.2 0.1415 65 2. 0 to 3.584 0 to 15.78 0 to 119.4 0 to 525.9 0.08371 80 3 0 to 5.429 0 to 23.91 0 to 180.9 0 to 796.7 0.05526 100 4 0 to 8.483 0 to 37.35 0 to 282.7 0 to 1244 0.03537 150 6 0 to 19.09 0 to 84.04 0 to 636.1 0 to 2800 0.01572 200 8 0 to 33.93 0 to 149.4 0 to 1130 0 to 4979 0.008842 2.94 MPa (426 psi) 1.96 MPa (284 psi) -0.098 MPa (-14.2 psi) -20 C (-4 F) 80 C (176 F) 100 C (212 F) 2.94 MPa (426 psi) -0.098 MPa (-14.2 psi) 0.98 MPa (142 psi) -20 C (-4 F) 80 C (176 F) 120 C (248 F) 130 C (266 F) |