No. SS2-MCB100-0100 Azbil Corporation - 2 - FUNCTIONAL SPECIFICATIONS Type of protection IEC IP65 (for indoor use only) Output signal Analog output 4 to 20 mA DC Load resistance: 0 to 400 Ω. Pulse output Select either pulse output or contact output. Pulse output (Totalized value output) Open collector Contact capacitance: 30V DC max., 100 mA max. Pulse width: 1 ms to 1 s Adjustable between 1ms to 1s or fixed at 50% of duty. Pulse frequency: 0.0001 to 500 Hz Contact output Open collector Contact capacitance: 30 V DC max., 100mA max. Alarm for Hi-Lo limit or failure status output Flow unit Volume flow: m3, L, cm3, G, mG, kG, B (barrel) Mass flow : t, kg, g, lb. Time : d, h, m, s Display Display: LCD Main display: 7-segment, 8 digits Sub display: 16 digits, 2 lines Display contents: Demonstrates three values simultaneously % flow rate, Actual flow rate, Totalized value Data setting Operation by four key switches Damping Adjustable between 0.5 and 199.9 seconds. Default setting: 5 seconds Low flow cutoff Adjustable between OFF (0%) and 10% of setting range. Below selected value, output is driven to the zero flow rate signal level. Default setting: OFF (0%) Dropout Adjustable between 0 and 10% of setting range. Below selected value, pulse output is cut. Default setting: 2% Empty pipe detection If the detector is empty, analog output and pulse output will be fixed at zero. The display will be latched to zero and “EMPTY” will be indicated. Failure diagnostics If the converter malfunctions, the analog output will be automatically fixed at 3.6 mA. Power failure An EEPROM retains data record of totalized flow volume when pulse output is used. Power supply 24V DC ± 10% or 100V AC ± 10% Power consumption 2.4W max. Ambient temperature limits 0 to 50°C Ambient humidity limits 5 to 85% RH (no condensation) Optional specifications Traceability certificate The following three documents are included. . Traceability system chart . Traceability certificate . Test report PHYSICAL SPECIFICATIONS Case material Polycarbonate Case cover material Polycarbonate INSTALLATION Cable gland Plastic gland (3 pieces) Cable Applicable cable outer diameter: φ6 to φ12 mm Applicable cable conductor Power supply: AWG14 to 22 (0.32 to 2.03 mm2) Outputs: AWG16 to 26 (0.13 to 1.31 mm2) Mounting Integral type Grounding Grounding resistance should be less than 100 Ω. Vibration effect Integral style: 4.9 m/s2 max. (0.5 G max.) For converter |