【DO!BOOK・ページリンク】 fi-04 197 / 228 10秒後にBOOKのページに移動します |
|
SS2-MVC300-0100 Azbil Corporation - 10 - MEASUREMENT RANGE SETTING Factory Default Setting (Unit:mm) Example: In the case of 50mm and 1 MPa Check the range of guaranteed accuracy in Table 1. The range of guaranteed accuracy is from 1,196 kg/h (upperlimit value) to 120 kg/h (lower-limit value). The range can be expanded to 1,255 kg/h which is 1.05 times as large as the upper-limit value of the range of guaranteed accuracy. It is recommended that you make an easy-touse setting* by omitting fractions, for example, 0 to 1,200 kg/h. If the lower-limit value is less than 69 kg/h, output will be cut. Use caution in this regard. * A special communicator is needed to change the internal setting. Figure 8 Measurement range and the range of guaranteed accuracy 0 50 100 105 (%) Lower-limit value of measurable range Measurable range Upper-limit value of measurable range Lower-limit value of the range of guaranteed accuracy (one tenth of the upper-limit value of measurable range) Range of guaranteed accuracy Upper-limit value of the range of guaranteed accuracy ITEM Factory Default Setting other Flow rate range Customer instruction Order decisions Process Pressure range 0.101 to 3.5 MPa_abs Temperature range 0 to 300°C Pulse Weigh Customer instruction Damping Customer instruction When there is not a designation this makes 2 seconds. Burnout Customer instruction Order decisions Low flow cut 3% of Flow rate range This converts Low flow cut, SP cut and DP cut to the flow rate SP cut value 0.035MPa and the biggest one is choose in it. DP cut value 0.3 kPa Density compensation Customer instruction Flow rate output : C only Height Customer instruction When there is not a designation this makes 0.0m.(MVC32/ 33A only) |